Skip to main content

Clinical application of the HM-1000 image processing for HER2 fluorescence in situ hybridization signal quantification in breast cancer

Abstract

Background

Accurate quantification of human epidermal growth factor receptor 2 (HER2) gene amplification is important for predicting treatment response and prognosis in patients with breast cancer. Fluorescence in situ hybridization (FISH) is the gold standard for the diagnosis of HER2 status, particularly in cases with equivocal status on immunohistochemistry (IHC) staining, but has some limitations of non-classical amplifications and such cases are diagnosed basing on additional IHC and FISH. This study investigated the clinical utility of a novel super-resolution fluorescence microscopy technique for the better FISH signal visualization and HER2 FISH classification.

Methods

Fourteen breast cancer tissue samples were retrospectively collected between September 2018 and February 2022, and FISH HER2 signal quantification was evaluated by determining the HER2/chromosome 17 centromere (CEP17) ratio and the number of HER2 signals per nucleus in super- versus conventional-resolution images.

Results

Super-resolution images maintained the same overall HER2 diagnosis from routine, but HER2 FISH amplification changed negative to monosomy in two cases. Two Letrozole non-response relapses coincided to monosomy samples. The median number of HER2 signals per nucleus was 7.5 in super-resolution images and 4.0 in conventional-resolution images in HER2-positive samples and 2.8 and 2.1 signals per nucleus, respectively, in HER2-negative samples.

Conclusions

Super-resolution images improved signal visualization, including a significant difference in the number of countable HER2 and CEP17 signals in a single nucleus compared with conventional-resolution images. Increased accuracy of signal quantification by super-resolution microscopy may provide clinicians with more detailed information regarding HER2 FISH status that allows to better FISH classification such as HER2-low samples.

Background

In patients with breast cancer, evaluation of human epidermal growth factor receptor 2 (HER2) expression is essential to determine whether targeted anti-HER2 treatment is appropriate. In routine clinical practice, HER2 expression status is tested by immunohistochemistry (IHC) and fluorescence in situ hybridization (FISH). Currently, FISH is more commonly used for evaluation of HER2 status and is now included as part of the standardized criteria for diagnosis in the 2018 American Society of Clinical Oncology (ASCO)/College of American Pathologists (CAP) HER2 testing guideline update [1]. Interpretation of FISH is based on quantification of HER2 and chromosome 17 centromere (CEP17) gene amplification signals [2]. Several studies have provided evidence of the correlation between HER2 amplification by FISH and response to anti-HER2 treatment [3,4,5]. These studies demonstrated that correct quantification of HER2 amplification, especially in samples with high HER2 expression, is important for predicting treatment response and patient prognosis in both the adjuvant and neoadjuvant setting.

Diagnosis of HER2 status based on FISH is the gold standard for cases with equivocal status on IHC staining, but there are some limitations associated with evaluation and signal quantification due to the low image quality and resolution [6]. When using a conventional fluorescence microscope, signals are sometimes blurred or overlapped with other signals. This can happen when the HER2 signals show an amplification cluster, the samples have high background noise and weak signals, or the signals at different depths are blurred or not visible at the same focus level. Ambiguity of HER2 signal images can cause inter-observer variability and uncountable or inaccurate results. In the past two decades, automated quantification of FISH signal amplification for HER2 status has been developed as an approach to solve some of these problems [7,8,9,10]. Different image analysis software and deep-learning systems have been established, but the fundamental limitation related to resolution and background noise due to autofluorescence in sample tissues remains unresolved.

Super-resolution microscopy is a series of technologies currently used in basic and advanced research, with some technologies achieving approximately 20-nm resolution [11]. Nanoreso (Sysmex, Kobe, Japan) is one such technology that captures the autonomous blinking of fluorophores in thousands of image frames and creates super-resolution fluorescent images based on Gaussian fitting of bright fluorescent spots. Nanoreso super-resolution microscopy normalizes the FISH signal without the need for additional staining and saves the captured images as a compatible digital file. This technology improves image resolution and provides a digitalized image for standardized quantification and signal localization that may be applicable for use in daily clinical practice.

Material and methods

The aim of this study was to compare HER2 signal quantification between conventional-resolution and super-resolution images using the HM-1000 (Sysmex, Kobe, Japan), a single-molecule fluorescence microscope with applied Nanoreso technology. The HM-1000 microscope may afford more precise quantification of HER2 amplification in patients with breast cancer, thereby providing better categorization of HER2 status for therapeutic decision making.

Retrospective breast cancer tissue samples with FISH results from routine practice were collected between September 2018 to February 2022 at the Anatomic Pathology Laboratory in Hospital Vall d’Hebron, Spain. Patient data were collected and anonymized before the analysis was performed. The samples were previously tested by IHC staining and FISH based on the 2018 ASCO/CAP HER2 testing guideline update [1]. HER2 FISH status was achieved by evaluating the HER2/CEP17 ratio and the number of HER2 signals per nucleus with super-resolution and conventional-resolution images.

FISH imaging

The collected tissue samples were de-paraffined and stained by dual-probe FISH for HER2 and CEP17 (Cytocell Ltd, Cambridge, UK) which the concentration of HER2 probe was 13.5 ng/μL and the CEP17 probe was 4 ng/μL, and by 0.125 µg/mL 4',6-diamidino-2-phenylindole (DAPI) for cell nuclei in 3-μm thick samples according to the following protocol.

Pre-treatment of samples was conducted using the Cytocell Tissue Pretreatment kit (RLPS100), in which the pre-treatment solution was applied for 30 minutes at 98°C and the samples underwent enzyme digestion with pepsin at 37°C for 20 minutes. Next, 15 μL of Cytocell HER2 (ERBB2) Amplification (LPS 001) probe was applied, and the samples were denatured at 75°C for 5 minutes and hybridized at 37°C overnight. After hybridization, the slides were washed with astringent solution 0.4× saline sodium citrate (SSC) solution (pH7.0) at 72°C for 2 minutes and 2× SSC + 0.005% Tween-20 solution (pH7.0) for 30 seconds.

Before FISH imaging, 15 μL of DAPI diluted in 1:5 by the imaging buffer for Oxygen scavenging were added to the samples, which were then each sealed with a cover glass.

Image digitalization

HER2 and CEP17 signals were captured using the HM-1000 fluorescence microscope with 100x/1.4 oil-immersion objective lens. Using 40% laser intensity, the excitation wave of each gene marker fluorochrome was 488 nm and 561 nm, respectively. The exposure time was 30 ms per frame, and 5000 frames per field were captured to reconstruct the super-resolution image to approximately 20-nm resolution. For nuclei imaging, the 405-nm excitation wave with 20% laser intensity was used and 1000 frames per field were captured. To select the optimal sample field, we utilized the 10 × 10 tiling mode of the HM-1000 software to obtain the image data for 100 fields. The captured images were automatically saved as TIFF files.

Digital image processing

The captured images were processed in four steps by machine learning and prediction techniques to reduce autofluorescence. First, the image file, which contained 5000 frames captured by the HM-1000, was separated into single image files using Python programming language. A total of 15,000 image files from three samples were used for Noise2Void model training, with the training images stored in TensorFlow [12, 13]. Sample images for the study were also separated in single image files and applied to the trained model to optimize the noise on each file using ImageJ software (U.S. National Institutes of Health, Bethesda, USA). Second, background noise was estimated using the rolling-ball algorithm plug-in and subtracted to obtain cleaned images (Castle and Keller [14]). Third, background thresholds were determined based on a linear regression model and machine learning. Finally, the separated 5000 image files were combined into one file using Python and adjusted by the slice-keeper function of ImageJ. Using ThunderSTORM, the modular plug-in for ImageJ that is used for sub-diffraction localization of molecules [15], Gaussian normalization was applied to each image, which consisted of 5000 noise-reduced frames, to create the super-resolution image. All developed scripts are available in the website GitHub [16].

Signal quantification

Fluorescence signals were counted manually by two observers. The nuclei with clustered or ambiguous signals were estimated from the distinguishable signal count or defined as more than four signals. HER2/CEP17 ratio was calculated from the number of nuclei in which CEP17 signals were countable. HER2-positive or -negative status for the images using conventional-resolution and super-resolution were based on the 2018 ASCO/CAP HER2 testing guideline update [1]. Conventional-resolution was defined as the image captured using the HM-1000 fluorescence microscope without digital image processing.

Statistical analysis

Commercialized software for statistical analysis, NCSS 10 (NCSS LLC., Kaysville, Utah, USA) was used for the analyses. Descriptive statistics, including mean, standard deviation (SD), median and range for continuous variables and frequency and percentage for categorical variables, were used to present the data. Normality tests were performed to define the statistical method for each significance test. Statistical significance (p-value) between two independent subgroups was determined using non-parametric tests: the Mann–Whitney U test was used for ordinal or continuous independent variables, the Chi-squared test for nominal independent variables, and the Wilcoxon test for paired samples. A p-value of <0.05 was considered statistically significant. The effect of each parameter was presented using a dot plot, percentage and 95% confidence intervals.

Results

Tissue samples

A total of 14 formalin-fixed, paraffin-embedded tissue samples were analyzed. The main characteristics of the patients in this study are summarized in Table 1. According to the routine hospital FISH-based diagnosis, nine samples (64.3%) were HER2 negative and five (35.7%) were HER2 positive. The overall HER2 diagnosis integrating IHC and FISH showed no difference between conventional-resolution image and super-resolution image using HM-1000–captured images. Demographic variables showed no significant differences between HER2-positive and HER2-negative patients. In HER2-positive samples, the median HER2 signals per nucleus was 7.5 in super-resolution images and 4.0 in conventional-resolution images. In HER2-negative samples, there was a median of 2.8 HER2 signals per nucleus in super-resolution images and 2.1 signals per nucleus in conventional-resolution images.

Table 1 Patient characteristics by HER2 status according to routine diagnosis

Noise optimization

As ThunderSTORM is very sensitive to background noise, noise reduction was applied for correct sample observation. The noise reduction process was performed using a machine learning model (Fig. 1). The original image included high background noise, which obstructed visualization of the HER2 signal. After application of the rolling-ball algorithm and the background threshold, significant noise reduction was observed. The background threshold for each image was estimated using the linear regression model (Fig. 2).

Fig. 1
figure 1

Noise optimization by machine learning model. Noise optimization and background subtraction for CEP17 signals, showing (a) the original image colored in 3-3-2 RGB; (b) the image after application of the rolling ball algorithm, where the radius pixels used for background subtraction is 15; and (c) after application of the threshold (1490), which was automatically obtained from the linear regression model

Fig. 2
figure 2

Linear regression model used to estimate the optimum background threshold. The preset linear regression model was used to determine the optimum threshold, then the reverse threshold was determined and the >95% correlation between the optimum threshold value and the mean (standard deviation) of the image was calculated. The optimum threshold value of each image was estimated using this regression model

Quantification of HER2 and CEP17 signals conventional-resolution and super-resolution images were compared using a single nucleus from HER2-positive and HER2-negative tissues (Fig. 3). In both conventional-resolution and super-resolution images, three HER2 signals and two CEP17 signals were observed in a single nucleus from HER2-negative tissue. In contrast, in a single nucleus of HER2-positive tissue, HER2 signals were not quantifiable and only one CEP17 signal was visible by conventional-resolution, whereas 12 HER2 signals and one CEP17 signal were observed using super-resolution. Table 2 describes the average HER2 and CEP17 signal count from the 14 tissue samples. In routine HER2-negative samples, total counted signals of HER2 and CEP17 had significant difference between conventional-resolution and super-resolution methods. In conventional-resolution images, there were three HER2-negative samples and five HER2-positive samples with ambiguous HER2 signals, and five HER2-negative samples and two HER2-positive samples with ambiguous CEP17 signals. In super-resolution images, all HER2 signals and CEP17 signals were countable in all samples.

Fig. 3
figure 3

Conventional-resolution vs super-resolution images of nucleus from HER2-positive and HER2-negative tissues. HER2 and CEP17 quantification and HER2 status evaluation in breast cancer tissue samples with dual-probe FISH for HER2 and CEP17 amplification, where red signals indicate HER2 and green signals indicate CEP17. Images show conventional-resolution of a single nucleus in (a) HER2-positive and (b) HER2-negative tissue, and super-resolution image of the nucleus of (c) HER2-positive and (d) HER2-negative tissue. The HER2 status was determined by routine FISH. The scale bar is 2 μm

Table 2 Comparison of HER2 signal quantification using dual-probe FISH between conventional-resolution and super-resolution image techniques

The counted signals classify HER2 FISH amplification in 5 groups basing on the criteria of 2018 ASCO/CAP HER2 testing guideline update (Table 3). Three samples were classified as group 2, previously known as monosomy by super-resolution images and two of them were classified as HER2 non-amplification from conventional-resolution images.

Table 3 Contingency table of super-resolution and conventional resolution FISH image evaluation

The overall HER2 status of such cases did not change because the group 2 samples were further examined by IHC and judged negative from IHC 1+ results.

There were three IHC 2+ samples included in this study; using routine FISH, two were diagnosed as HER2 negative and one was diagnosed as HER2 positive. The IHC category and HER2 signals quantified by super-resolution images were correlated (Fig. 4).

Fig. 4
figure 4

Dot plot of HER2 FISH signals quantified by conventional-resolution and super-resolution. Conventional-resolution (red dots) and super-resolution (blue dots). The HER2 signals per nucleus were calculated by dividing the number of HER2 signals by the total number of nuclei counted in each image

Patient follow-up and FISH status

Two of 14 patients did not respond Letrozole treatment and relapsed later. The comparison between clinical follow-up data and HER2 FISH amplification demonstrated that these two samples were classified as group 2 using super-resolution images. One of them were classified as group 2 by super-resolution image but as HER2 non-amplification by conventional-resolution image (Fig. 5).

Fig. 5
figure 5

Conventional-resolution vs super-resolution images from Letrozole non-responded relapse cases. Red signals indicate HER2 and green signals indicate CEP17. Images show the case classified as Group 2 by both resolution in (1) conventional-resolution and (2) super-resolution, and the case classified as group 2 by super-resolution image but as HER2 non-amplification by conventional-resolution image in (3) conventional-resolution and (4) super-resolution. The signals in super-resolution are much sharper than conventional-resolution. The scale bar is 10 μm

Discussion

During the decade after the 2013 ASCO/CAP guideline was published [17], there was a lack of clear guidance around how to diagnose non-classical HER2 FISH patterns. The 2018 ASCO/CAP HER2 testing guideline update provided a clear guidance of Group 1 to 5 diagnosis [1], which indicates additional IHC and FISH evaluation for non-classical HER2 FISH patterns. On the other hand, the correct quantification of FISH signals and the solution of further improvement of FISH images were not tackled so much. In this study, we aimed to use super-resolution fluorescence microscopy for HER2 signal quantification, with an expectation of contributing better evaluation of HER2 status.

This study showed a significant difference in HER2 signal quantification between using conventional-resolution and super-resolution images from the HM-1000 fluorescence microscope. Super-resolution images provided better visualization of blurred and overlapped FISH signals, which are particularly common in HER2-positive samples. When evaluating CEP17 signals, some samples showed faint signals that were difficult to identify in conventional-resolution images but were countable in super-resolution images. For this reason, there was a significant difference between conventional- and super-resolution images in the number of nuclei with countable CEP17 signals. The better visualization was the result of the improvement of resolution, as well as a reduction in autofluorescence. Some samples can have high levels of autofluorescence and background noise, which can lead to confusion of the probe-specific and non-specific signals. In addition to the original function of HM1000, the Noise2Void and rolling-ball algorithms, together with machine learning, reduced the effect of autofluorescence and corrected the defocused signals in the captured images, while the ThunderSTORM plug-in acquired super-resolution images using the point-spread function of the signals from the individual molecules.

The improvement in resolution may also have a role in the accurate detection of centromeric copy number, which appears to be an important factor in chromosomal instability (CIN). In studies of breast cancer samples [18] or multiple tumor types [19], gains and losses in DNA were reported to lead to CIN and aneuploidy. The super-resolution method of HER2 quantification could be useful in detecting CIN, which seem to be associated with poor prognosis and low treatment response in patients with breast cancer [18].

We observed that in HER2-positive samples, there was a significant difference in the number of counted HER2 signals per nucleus between conventional- and super-resolution images, but according to the current HER2 classification, all samples with >4 HER2 signals are considered positive, with no further classification based on the quantity of HER2 signals above this level. However, other studies have indicated that high levels of HER2 amplification may predict pathological complete response (pCR) to anti-HER2 therapy in patients with breast cancer. A study by Singer and colleagues showed that early breast cancer patients with a HER2/CEP17 ratio of >6 had significantly higher rates of pCR to neoadjuvant trastuzumab treatment than those with low HER2 amplification levels [20]. In a similar study by Choi and colleagues, breast cancer patients with pCR to neoadjuvant trastuzumab + pertuzumab treatment had a median HER2/CEP17 ratio of 7.08 and a median HER2 copy number of 17, whereas patients without pCR had a median ratio of 4.70 and a median copy number of 12 [21]. Antolín and colleagues subsequently demonstrated a significant direct correlation between pCR to neoadjuvant chemotherapy + trastuzumab and high HER2 amplification, with pCR rates of 65% in patients with a HER2/CEP17 ratio of >5 and 61% in those with >10 HER2 signals per nucleus [3].

Also, the better quantification by super-resolution images did not change the final HER2 status in this study because the samples which are not clearly positive or negative with observing FISH amplification, so-called non-classical HER2 amplification were examined by IHC and judged by IHC result. We observed two samples which changed their HER2 FISH classification from negative to monosomy which is non-classical HER2 amplification by super-resolution method. In future investigation, we would like to evaluate significant number of samples to confirm such FISH classification change.

This result suggests that super-resolution method may improve the differentiation between HER2-low and HER2-negative to better categorize the patients who can benefit from new treatments that target HER2. This study faced two Letrozole non-responded relapse cases which coincide group 2 according to the 2018 ASCO/CAP definition. This could provide a spark for HER2-low definition to correlate the clinical outcome. There are several ongoing studies in patients with HER2-low breast cancer, several of which are exploring the minimum HER2 expression threshold required for drug efficacy [22]. In this regard, a more precise study of the impact of gene copy number is also warranted.

Several studies have examined whether automated quantification can improve the inter-observer variability of HER2 [23] and Ki67 [24] amplification in breast cancer samples. These studies both used machine-learning technology for quantitative digital image analysis that resulted in excellent inter-observer reproducibility and concordance with pathologist assessment [23, 24]. In future studies, we plan to evaluate whether we can apply similar technology to automatically quantify HER2 amplification in breast cancer patients using super-resolution HM-1000 images.

This study had some technical limitations. The routine sample preparation protocol in hospital and the de-paraffination process may have caused high levels of background noise in some samples. The HM-1000 fluorescence microscope builds in four excitation lasers but does not have emission or excitation light filters, so noise-free images could not be obtained. This problem may be resolved by using a specific fluorochrome that considers the excitation wave, but commercially available in vitro diagnostic probes were used in this study considering to apply super-resolution method in future clinical practice with less impact for the actual routine. Machine learning and prediction techniques were utilized to reduce background noise, but these methods should be validated in future studies if the processed image only contains HER2- or CEP17-specific signals. Although the small sample size, we found statistically significant difference in the signal count for diagnosis. The prospective validation study with bigger rate of tumors in equivocal status could confirm the clinical application. 

Conclusions

Overall HER2 status evaluated by conventional-resolution and by super-resolution images maintained the same; however, HER2 FISH amplification patterns changed from negative to monosomy, and the number of counted HER2 signals in HER2-positive samples increased with super-resolution versus conventional-resolution images. Improvements in the accuracy of HER2 signal quantification may give clinicians further information regarding HER2 expression status and potentially allow for improved precision with regard to therapeutic decisions especially in patients with HER2-low breast cancer.

Availability of data and materials

The data that support the findings of this study are available within the paper. Raw data files are not openly available due to reasons of sensitivity and confidentiality.

Abbreviations

HER2:

Human epidermal growth factor receptor 2

FISH:

Fluorescence in situ hybridization

IHC:

Immunohistochemistry

CEP17:

Chromosome 17 centromere

ASCO:

American Society of Clinical Oncology

CAP:

College of American Pathologists

DAPI:

4',6-diamidino-2-phenylindole

SSC:

Saline sodium citrate

SD:

Standard deviation

CIN:

Chromosomal instability

pCR:

Pathological complete response

References

  1. Wolff AC, Hammond MEH, Allison KH, Harvey BE, Mangu PB, Bartlett JMS, et al. Human epidermal growth factor receptor 2 testing in breast cancer: American Society of Clinical Oncology/College of American Pathologists clinical practice guideline focused update. J Clin Oncol. 2018;36(20):2105–22.

    Article  CAS  PubMed  Google Scholar 

  2. Hanna WM, Rüschoff J, Bilous M, Coudry RA, Dowsett M, Osamura RY, et al. HER2 in situ hybridization in breast cancer: clinical implications of polysomy 17 and genetic heterogeneity. Mod Pathol. 2014;27(1):4–18.

    Article  CAS  PubMed  Google Scholar 

  3. Antolín S, García-Caballero L, Reboredo C, Molina A, Mosquera J, Vázquez-Boquete Á, et al. Is there a correlation between HER2 gene amplification level and response to neoadjuvant treatment with trastuzumab and chemotherapy in HER2-positive breast cancer? Virchows Arch. 2021;479(4):853–7.

    Article  PubMed  Google Scholar 

  4. Borley A, Mercer T, Morgan M, Dutton P, Barrett-Lee P, Brunelli M, et al. Impact of HER2 copy number in IHC2+/FISH-amplified breast cancer on outcome of adjuvant trastuzumab treatment in a large UK cancer network. Br J Cancer. 2014;110(8):2139–43.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. Wu Z, Xu S, Zhou L, Yin W, Lin Y, Du Y, et al. Clinical significance of quantitative HER2 gene amplification as related to its predictive value in breast cancer patients in neoadjuvant setting. Onco Targets Ther. 2018;11:801–8.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Gu J, School of Health Professions, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030, USA, Tang Z, Chen H, Sfamenos S, Geiersbach KB, et al. HER2 FISH for breast cancer: Advances in quantitative image analysis and automation. OBM Genet. 2020;4(2):1–14.

    Article  Google Scholar 

  7. Furrer D, Jacob S, Caron C, Sanschagrin F, Provencher L, Diorio C. Validation of a new classifier for the automated analysis of the human epidermal growth factor receptor 2 (HER2) gene amplification in breast cancer specimens. Diagn Pathol. 2013;8(1):17.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Höfener H, Homeyer A, Förster M, Drieschner N, Schildhaus H-U, Hahn HK. Automated density-based counting of FISH amplification signals for HER2 status assessment. Comput Methods Programs Biomed. 2019;173:77–85.

    Article  PubMed  Google Scholar 

  9. Klijanienko J, Couturier J, Galut M, El-Naggar AK, Maciorowski Z, Padoy E, et al. Detection and quantitation by fluorescence in situ hybridization (FISH) and image analysis of HER-2/neu gene amplification in breast cancer fine-needle samples. Cancer. 1999;87(5):312–8.

    Article  CAS  PubMed  Google Scholar 

  10. Konsti J, Lundin J, Jumppanen M, Lundin M, Viitanen A, Isola J. A public-domain image processing tool for automated quantification of fluorescence in situ hybridisation signals. J Clin Pathol. 2008;61(3):278–82.

    Article  CAS  PubMed  Google Scholar 

  11. Schermelleh L, Ferrand A, Huser T, Eggeling C, Sauer M, Biehlmaier O, et al. Super-resolution microscopy demystified. Nat Cell Biol. 2019;21(1):72–84.

    Article  CAS  PubMed  Google Scholar 

  12. Abadi M, Agarwal A, Barham P, Brevdo E, Chen Z, Citro C, et al. TensorFlow: Large-scale machine learning on heterogeneous distributed systems. 2016. https://0-doi-org.brum.beds.ac.uk/10.48550/ARXIV.1603.04467. Accessed 14 March 2016.

  13. Krull A, Buchholz T-O, Jug F. Noise2Void - learning denoising from single noisy images. 2018. https://0-doi-org.brum.beds.ac.uk/10.48550/ARXIV.1811.10980. Accessed 5 April 2019

  14. Castle M, Keller J, Schmid M. Rolling ball background subtraction (ImageJ). 2007. https://imagej.net/plugins/rolling-ball-background-subtraction. Accessed 6 Feb 2024.

  15. Ovesný M, Křížek P, Borkovec J, Svindrych Z, Hagen GM. ThunderSTORM: a comprehensive ImageJ plug-in for PALM and STORM data analysis and super-resolution imaging. Bioinformatics. 2014;30(16):2389–90.

    Article  PubMed  PubMed Central  Google Scholar 

  16. https://github.com/SysmexRDCE/HM1000.git [Accessed on January 8, 2024]

  17. Wolff AC, Hammond MEH, Hicks DG, Dowsett M, McShane LM, Allison KH, et al. Recommendations for human epidermal growth factor receptor 2 testing in breast cancer: American Society of Clinical Oncology/College of American Pathologists clinical practice guideline update. Arch Pathol Lab Med. 2014;138(2):241–56.

    Article  PubMed  Google Scholar 

  18. Lee K, Kim HJ, Jang MH, Lee S, Ahn S, Park SY. Centromere 17 copy number gain reflects chromosomal instability in breast cancer. Sci Rep. 2019;9(1):17968.

    Article  ADS  PubMed  PubMed Central  Google Scholar 

  19. Carter SL, Eklund AC, Kohane IS, Harris LN, Szallasi Z. A signature of chromosomal instability inferred from gene expression profiles predicts clinical outcome in multiple human cancers. Nat Genet. 2006;38(9):1043–8.

    Article  CAS  PubMed  Google Scholar 

  20. Singer CF, Tan YY, Fitzal F, Steger GG, Egle D, Reiner A, et al. Pathological complete response to neoadjuvant trastuzumab is dependent on HER2/CEP17 ratio in HER2-amplified early breast cancer. Clin Cancer Res. 2017;23(14):3676–83.

    Article  CAS  PubMed  Google Scholar 

  21. Choi JH, Jeon CW, Kim YO, Jung S. Pathological complete response to neoadjuvant trastuzumab and pertuzumab therapy is related to human epidermal growth factor receptor 2 (HER2) amplification level in HER2-amplified breast cancer. Medicine (Baltimore). 2020;99(46):e23053.

    Article  CAS  PubMed  Google Scholar 

  22. Modi S, Jacot W, Yamashita T, Sohn J, Vidal M, Tokunaga E, et al. Trastuzumab deruxtecan in previously treated HER2-low advanced breast cancer. N Engl J Med. 2022;387(1):9–20.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Turashvili G, Leung S, Turbin D, Montgomery K, Gilks B, West R, et al. Inter-observer reproducibility of HER2 immunohistochemical assessment and concordance with fluorescent in situ hybridization (FISH): pathologist assessment compared to quantitative image analysis. BMC Cancer. 2009;9(1):165.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Acs B, Pelekanou V, Bai Y, Martinez-Morilla S, Toki M, Leung SCY, et al. Ki67 reproducibility using digital image analysis: an inter-platform and inter-operator study. Lab Invest. 2019;99(1):107–17.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was supported by Sysmex R&D Center Europe related to reagents and materials.

Author information

Authors and Affiliations

Authors

Contributions

V.P. study design, samples evaluation, statistical analysis, manuscript elaboration; T.M. ISH probes, read and approved manuscripts drafts, M.R. image analysis, read and approved manuscript, Y.S. samples evaluation, statistical analysis, manuscript elaboration; S.RyC. manuscript elaboration.

Corresponding author

Correspondence to Vicente Peg.

Ethics declarations

Ethics approval and consent to participate

The study was carried out without additional intervention of patients, and all tissue samples were processed under the protocol approved by the ethical committee of Vall d’Hebron Hospital. Patient data used for this study were anonymized before the analysis was conducted.

Consent for publication

Not applicable.

Competing interests

V.P. has received fees as consultant, participated in advisory boards and received travel grants from Sysmex, Roche, Merck Sharp & Dohme, AstraZeneca, Bayer, Exact Sciences and Daiichi Sankyo. Y.S. is an employee of Sysmex R&D Center Europe GmbH.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Peg, V., Moline, T., Roig, M. et al. Clinical application of the HM-1000 image processing for HER2 fluorescence in situ hybridization signal quantification in breast cancer. Diagn Pathol 19, 32 (2024). https://0-doi-org.brum.beds.ac.uk/10.1186/s13000-024-01455-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s13000-024-01455-8

Keywords